Emergency tourniquet
Emergency tourniquets are cuff-like devices designed to stop severe traumatic bleeding before or during transport to a care facility. They are wrapped around the limb, proximal to the site of trauma, and tightened until all blood vessels underneath are occluded. The design and construction of emergency tourniquets allows quick application by first aid responders or the injured persons themselves.[1] Correct use of tourniquet devices has been shown to save lives under austere conditions with comparatively low risk of injury. In field trials, prompt application of emergency tourniquets before the patient goes into shock are associated with higher survival rates than any other scenario where tourniquets were used later or not at all.[2][3]
Tourniquet design
[edit]Existing guidelines call for the use of improvised "rope-and-stick" tourniquets as a last resort to stop severe bleeding. However, purpose-made tourniquet devices that are well designed can provide greatly increased safety and efficacy.[2][4] Variability in performance has been shown to exist between various designs and application methods.[5][6]
Mechanical advantage
[edit]Mechanisms that confer sufficient mechanical advantage are essential for applying adequate pressure to stop bleeding, particularly on the lower extremities. Pressures that occlude venous but not arterial flow can exacerbate hemorrhage and cause damage to healthy tissue.[1]
Mechanical characteristics of emergency tourniquet devices
[edit]Tourniquet | Strap width (cm) [1] | Mechanism [1] | Note [1] |
---|---|---|---|
SAM XT Tourniquet | 3.7 | Windlass and Buckle | Buckle sets with correct force; windlass finishes pressure |
Combat Application Tourniquet (CAT) | 3.8 | Windlass | Stick and strap inside outer sleeve |
Emergency & Military Tourniquet (EMT) | 9.1 | Pneumatic | Hand pump and inflatable bladder |
K2 Tactical Tourniquet (K2) | 3.8 | Clamp | Modified wood clamp |
Smart Tactical Application Tourniquet (S.T.A.T.) | 2.5 | Ratchet | Strap with ratcheting mechanism |
Last Resort Tourniquet (LRT) | 5.1 | Ratchet | Strap with ratcheting mechanism |
London Bridge Tourniquet (LBT) | 2.4 | Ratchet | Strap with ratcheting mechanism |
Mechanical Advantage Tourniquet (MAT) | 3.8 | Block and tackle | Pulleys on outer frame |
One-Handed Tourniquet (OHT) | 2.8 | Elastic | Parallel bungee cords and clamp |
Self-Applied Tourniquet System (SATS) | 3.8 | cam | Cantilever system |
Special Operations Forces Tactical Tourniquet (SOFTT) | 3.7 | Windlass | Stick and strap |
Glia tourniquet | Variable | Windlass | Stick and strap |
Most commercial tourniquets cost in the range from $30-$50 (USD). Results from laboratory and field testing suggest that windlass and pneumatic mechanisms are effective where other systems fail due to excessive pain, slipping, inadequate force, or mechanical failure.[1]
Pressure gradients
[edit]Pressure underneath a tourniquet cuff is not evenly distributed, with the highest pressures localized around the cuff center line and decreasing to zero near the cuff edges.[7] A high rate of change of pressure across the cuff width, or a high cuff pressure gradient, is a leading cause of nerve and muscle injury from tourniquet use.[7] Tourniquets with wider straps or cuffs, especially those with pneumatic actuation in contrast to mechanical force, distribute pressure more evenly and produce lower pressure gradients.[7] They are therefore more likely to stop bleeding and less likely to cause damage to underlying tissue, in addition to being significantly less painful than tourniquets with narrow straps and bands.[4][8] Over pressure protection in certain emergency tourniquets also help to prevent excessive force from damaging the limb.[1]
Risks
[edit]Possible risks of complications—morbidity—related to emergency tourniquet use include
|
Emergency care services implementing routine tourniquet use, especially in the civilian setting, should exercise caution and ensure that training is adequate for optimal results.[3] However, given proper precautions, the occurrence of complications due to tourniquet use is quite rare.[9] Designed tourniquet devices are routinely tightened over healthy limbs during training with no ill effects, and recent evidence from combat hospitals in Iraq suggests that morbidity rates are low when users adhere to standard best practices. Since no better alternatives exist for users to self-apply with only basic training, the benefit of tourniquet use far outweighs the risks.[2][3][9]
Safe tourniquet practice involves:
1. Careful placement of tourniquet proximal to all sites of hemorrhage. 2. Limiting tourniquet time to less than two hours, if possible. 3. Minimizing excessive applied pressure beyond the point of complete blood flow cessation.[2][9] |
Current developments
[edit]Field experience
[edit]Latest field trials suggest that wider straps are more effective and less painful than tourniquets with thinner straps. The concept of limb occlusion pressure is also gaining prominence over the misconception that greater applied force results in greater effectiveness. In addition, studies of failed cases indicate that the correct devices should be coupled with training that facilitates realistic expectations and correct user actions.[4] The Stop The Bleed educational initiative provides knowledge aimed at the greater public on when to use a tourniquet and the correct user actions.[10]
Emerging needs
[edit]Despite the success of widespread tourniquet deployment to limit combat casualties, many preventable deaths from hemorrhage occur where conventional tourniquet use is inappropriate. The need exists for controlling junctional bleeding, especially in the pelvic area.[11] In 2012, the Combat Ready Clamp (CRoC) was selected by the U.S. Army Institute of Surgical Research (USAISR) for that purpose.[12] Another emerging need is more refined training regimes and doctrine based on scientific evidence, which can ensure that future tourniquet practice and policies are in line with the most current body of knowledge.[11]
See also
[edit]References
[edit]- ^ a b c d e f g Walters TJ, Wenke JC, Kauvar DS, McManus JG, Holcomb JB, Baer DG (2005). "Effectiveness of self-applied tourniquets in human volunteers". Prehospital Emergency Care. 9 (4). Informa Healthcare: 416–422. doi:10.1080/10903120500255123. PMID 16263675. S2CID 23392555.
- ^ a b c d e Kragh JF, Walters TJ, Baer DG, Fox CJ, Wade CE, Salinas J, Holcomb JB (February 2008). "Practical use of emergency tourniquets to stop bleeding in major limb trauma". The Journal of Trauma. 64 (2 Suppl). Lippincott Williams & Wilkins: S38–S50. doi:10.1097/TA.0b013e31816086b1. PMID 18376170.
- ^ a b c Kragh JF, Walters TJ, Baer DG, Fox CJ, Wade CE, Salinas J, Holcomb JB (January 2009). "Survival with emergency tourniquet use to stop bleeding in major limb trauma". Annals of Surgery. 249 (1). Lippincott Williams & Wilkins: 1–7. doi:10.1097/SLA.0b013e31818842ba. PMID 19106667. S2CID 42907438.
- ^ a b c Kragh JF, O'Neill ML, Walters TJ, Dubick MA, Baer DG, Wade CE, et al. (October 2011). "The military emergency tourniquet program's lessons learned with devices and designs". Military Medicine. 176 (10). Association of Military Surgeons of the U.S.: 1144–1152. doi:10.7205/milmed-d-11-00114. PMID 22128650.
- ^ Ruterbusch, VL; Swiergosz, MJ; Montgomery, LD; Hopper, KW; Gerth, WA (2005). "ONR/MARCORSYSCOM Evaluation of Self-Applied Tourniquets for Combat Applications". United States Navy Experimental Diving Unit Technical Report. NEDU-TR-05-15. Archived from the original on 2009-09-23. Retrieved 2008-07-22.
{{cite journal}}
: CS1 maint: unfit URL (link) - ^ Hill, JP; Montgomery, LD; Hopper, KW; Roy, LA (2007). "Evaluation of Self-Applied Tourniquets for Combat Applications, Second Phase". US Navy Experimental Diving Unit Technical Report. NEDU-TR-07-07. Archived from the original on 2009-09-21. Retrieved 2008-07-22.
{{cite journal}}
: CS1 maint: unfit URL (link) - ^ a b c Noordin S, McEwen JA, Kragh JF, Eisen A, Masri BA (December 2009). "Surgical tourniquets in orthopaedics". The Journal of Bone and Joint Surgery. American Volume. 91 (12). JBJS: 2958–2967. doi:10.2106/JBJS.I.00634. PMID 19952261.
- ^ Taylor DM, Vater GM, Parker PJ (September 2011). "An evaluation of two tourniquet systems for the control of prehospital lower limb hemorrhage". The Journal of Trauma. 71 (3). Lippincott Williams & Wilkins: 591–595. doi:10.1097/TA.0b013e31820e0e41. PMID 21768905.
- ^ a b c Kragh JF, Littrel ML, Jones JA, Walters TJ, Baer DG, Wade CE, Holcomb JB (December 2011). "Battle casualty survival with emergency tourniquet use to stop limb bleeding". The Journal of Emergency Medicine. 41 (6). Lippincott Williams & Wilkins: 590–597. doi:10.1016/j.jemermed.2009.07.022. PMID 19717268.
- ^ Jacobs L, Keating JJ, Hunt RC, Butler FK, Pons PT, Gestring M, et al. (October 2022). "Stop the BleedⓇ". Current Problems in Surgery. 59 (10): 101193. doi:10.1016/j.cpsurg.2022.101193. PMID 36253022. S2CID 251433574.
- ^ a b Murphy C, Kragh JF, Dubick MA (2011). "New Tourniquet Device Concepts for Battlefield Hemorrhage Control". U.S. Army Medical Department Journal: 38–48. ISSN 1524-0436. PMID 21607905.
- ^ "New truncal tourniquet ready for battle use".